5-27
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Service Instructions
Chamber Transducer
Probe Failure,
Continued
STATUS..
SERVICE INFORMATION:
212
CHAMBER TEMP PROBE FAILURE
→
RTD PROBE, CT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE CONTROL BOARD*
→
RECALIBRATE
PAPER
EXIT
FEED
For Healthcare
Only
*Depending upon the configura-
tion of the unit, screen text
may reference 146656-185
control board, or 146659-065
control board.
STATUS..
SERVICE INFORMATION:
212
CHAMBER TEMP PROBE FAILURE
→
RTD PROBE, CT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
RECALIBRATE
For Life Sciences
Only
PAPER
SERVICE
EXIT
FEED
MODE