5-17
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Operator Instructions
Occurs if door switch
does not make contact
within allotted time.
Too Long To Close Door
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
239
TOO LONG TO CLOSE DOOR
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION
UNTIL DOOR IS CLOSED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK DOOR FOR OBSTRUCTION
→
REMOVE OBSTRUCTION AND
CLOSE DOOR
3. IF DOOR WILL NOT CLOSE, CALL
SERVICE
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
240
TOO LONG TO CLOSE DOOR
→
DOOR SWITCH DID NOT MAKE IN
ALLOTTED TIME
CAUSES AND CORRECTION:
1.
DOOR SWITCH MALFUNCTION
→
CHECK LS1(LS2) CONNECTIONS
→
READJUST LS1(LS2)
→
REPAIR LS1(LS2)
2.
POWER DOOR MECHANISM FAILURE
→
REPAIR MECHANISM
→
REPLACE MOTOR
PAPER
CLEAR
EXIT
FEED
ALARM
The following alarm screens will appear only when the sterilizer is
not
processing
a cycle.
5.3 Out-of-Cycle
Alarms