5-20
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Operator Instructions
Waste Temperature
Probe Failure
Occurs if waste line
temperature reading is
outside the normal
range.
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM! WASTE
215
TEMPERATURE PROBE FAILURE
CHAMBER:
000.0 C
00.0 psig
WASTE:
000.0 C
STERILIZER WILL:
•
PREVENT NEW CYCLE FROM BEING
STARTED UNTIL ALARM IS CLEARED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
216
WASTE TEMP PROBE FAILURE
→
RTD PROBE, WT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
STATUS..
SERVICE INFORMATION:
216
WASTE TEMP PROBE FAILURE
→
RTD PROBE, WT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE CONTROL BOARD*
→
RECALIBRATE
PAPER
EXIT
FEED
For Healthcare
Only
*Depending upon the configura-
tion of the unit, screen text
may reference 146656-185
control board, or 146659-065
control board.