5-3
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Operator Instructions
Too Long In Charge,
Continued
Screen with Service Instructions
For Life Sciences
Only
STATUS
PAPER
FEED
SILENCE
ABORT
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
→
IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
205
TOO LONG IN CHARGE
For Healthcare
Only
→
CHAMBER DID NOT REACH STERILIZE
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 PSIG
→
CHECK STEAM SUPPLY PIPING
2.
STEAM REGULATOR MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
206
TOO LONG IN CHARGE
ABORT
PAPER
EXIT
FEED
For Healthcare
Only
→
CHAMBER DID NOT REACH STERILIZE
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 PSIG
→
CHECK STEAM SUPPLY PIPING
2.
STEAM REGULATOR MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
206
TOO LONG IN CHARGE
SERVICE
ABORT
EXIT
MODE