5-28
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Operator Instructions
Screen with Service Instructions
Jacket Temperature
Probe Failure
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
00.0 psig
JACKET:
000.0 C
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS..
ALARM! JACKET
213
TEMPERATURE PROBE FAILURE
Occurs if jacket tem-
perature reading is
outside the normal
range.
STATUS..
SERVICE INFORMATION:
214
JACKET TEMP PROBE FAILURE
→
RTD PROBE, JT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
STATUS..
SERVICE INFORMATION:
214
JACKET TEMP PROBE FAILURE
→
RTD PROBE, JT, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
REPAIR
2.
PROBE FAILED
→
REPLACE
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE CONTROL BOARD*
→
RECALIBRATE
PAPER
EXIT
FEED