5-11
Troubleshooting
Routine Maintenance
129367-410
Alarm
Description
Screen with Service Instructions
Occurs if chamber
pressure is 1.0 psig
below the fixed rate
during slow exhaust
phase. (Sterilizers with
Liquid cycle option
only.)
Exhaust Rate Too Slow
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
243
EXHAUST RATE TOO SLOW
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
TRY TO EXHAUST CHAMBER ACCORDING
TO OPTIMAL COOLING RATE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
Screen with Operator Instructions
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
234
TOO LONG IN STERILIZE
→
THE TIME SPENT IN THE
STERILIZE PHASE WAS TOO LONG
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
REPAIR S02
2.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S40
4.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
5.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE
STATUS..
SERVICE INFORMATION:
244
EXHAUST RATE TOO SLOW
PAPER
SERVICE
EXIT
FEED
MODE
→
CHAMBER EXHAUSTED SLOWERTHAN
THAN THE EXPECTED RATE
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
REPAIR S40
2.
SOLENOID VALVE MALFUNCTION
→
REPAIR S03
3.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
Too Long In Sterilize,
Continued