5-7
Troubleshooting
Routine Maintenance
129367-410
Occurs if chamber
does not air break the
vacuum to 2.0 inHg
within the allotted time.
Too Long In Air Break
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
225
TOO LONG IN AIR BREAK
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
EXTEND AIR BREAK TIME
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
Alarm
Description
Screen with Operator Instructions
STATUS..
SERVICE INFORMATION:
226
TOO LONG IN AIR BREAK
→
CHAMBER DID NOT AIR BREAK VACUUM
TO 2 inHg WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
AIR INLET FILTER PLUGGED
→
REPLACE
2.
SOLENOID VALVE MALFUNCTION
→
REPAIR S01
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
For Healthcare
Only
STATUS..
SERVICE INFORMATION:
226
TOO LONG IN AIR BREAK
→
CHAMBER DID NOT AIR BREAK VACUUM
TO 2 inHg WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
AIR INLET FILTER PLUGGED
→
REPLACE
2.
SOLENOID VALVE MALFUNCTION
→
REPAIR S01
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
EXIT
FEED
Screen with Service Instructions