5-4
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Operator Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
EXTEND EXHAUST TIME
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
STATUS..
ALARM!
200
TOO LONG IN EXHAUST
Occurs if chamber does
not exhaust to 4.0 psig
within the allotted time.
Too Long In Exhaust
→
CHAMBER DID NOT EXHAUST TO
ATMOSPHERIC PRESSURE WITHIN ALLOTTED
TIME
CAUSES AND CORRECTION:
1.
CHAMBER DRAIN STRAINER PLUGGED
→
CLEAN
2.
SOLENOID VALVE MALFUNCTION
→
REPAIR S03
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
201
TOO LONG IN EXHAUST
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
→
CHAMBER DID NOT EXHAUST TO
ATMOSPHERIC PRESSURE WITHIN ALLOTTED
TIME
CAUSES AND CORRECTION:
1.
CHAMBER DRAIN STRAINER PLUGGED
→
CLEAN
2.
SOLENOID VALVE MALFUNCTION
→
REPAIR S03
3.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
STATUS..
SERVICE INFORMATION:
201
TOO LONG IN EXHAUST
PAPER
EXIT
FEED
For Healthcare
Only
Screen with Service Instructions