Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Service Instructions
Too Long To Seal Door,
Continued
STATUS..
SERVICE INFORMATION:
228
TOO LONG TO SEAL DOOR
→
DOOR SEAL DID NOT REACH 10 PSIG
WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
READJUST PS1(PS2)
→
REPAIR PS1(PS2)
2.
SEAL NOT ACTIVATING
→
CHECK SEAL STEAM
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S35(S36)
ABORT
CLEAR
PAPER
EXIT
ALARM
FEED
For Healthcare
Only
Occurs if door seal
pressure does not
drop below 10 psig
within allotted time.
Too Long To Unseal
Door
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
229
TOO LONG TO UNSEAL DOOR
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS UNSEALED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
3. IF LOAD MUST BE REMOVED, REFER TO
EMERGENCY DOOR OPERATION PROCEDURE
IN OPERATING MANUAL.
STATUS..
SERVICE INFORMATION:
230
TOO LONG TO UNSEAL DOOR
→
DOOR SEAL PRESSURE NOT BELOW 10 PSIG
WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
READJUST PS1(PS2)
→
REPAIR PS1(PS2)
2.
SEAL NOT RETRACTING
→
CHECK SEAL STEAM
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S37(S38)
→
REPAIR S35(S36)
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
Screen with Operator Instructions
Screen with Service Instructions