5-8
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Operator Instructions
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
224
UNDER STERILIZE TEMPERATURE
→
CHAMBER TEMPERATURE DROPPED BELOW
STERILIZE TEMPERATURE BY UNDERTEMP
TEMPERATURE VALUE
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 PSIG
→
CHECK STEAM SUPPLY PIPING
2.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
REPAIR S09
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE
NOTE: If under-
temperature value is
set to "RESUME", the
sterilizer will auto-
matically resume
sterilize time after set
temperature is
reached.
For Life Sciences
Only
Occurs if chamber
temperature drops
below sterilize
temperature.
Under Sterilize
Temperature
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
223
UNDER STERILIZE TEMPERATURE
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY RESTART STERILIZE TIME
AND CONTINUE CYCLE AFTER SET
TEMP. IS REACHED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
For Life Sciences
Only
STATUS..
ALARM!
223
UNDER STERILIZE TEMPERATURE
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY RESTART STERILIZE TIME
AND CONTINUE CYCLE AFTER SET
TEMP. IS REACHED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
For Healthcare
Only
SILENCE
ABORT
SERVICE
ALARM
HELP
STATUS
PAPER
FEED