5-19
Troubleshooting
Routine Maintenance
129367-410
Pressure In Chamber
Occurs if 2.0 psig
pressure is sensed in
the chamber.
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
221
PRESSURE IN CHAMBER
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY EXHAUST CHAMBER TO
ATMOSPHERIC PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
Alarm
Description
Screen with Operator Instructions
STATUS..
SERVICE INFORMATION:
222
PRESSURE IN CHAMBER
→
2 PSIG PRESSURE SENSED IN CHAMBER
WHEN NOT IN CYCLE
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
REPAIR S02
2.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3.
TRANSDUCER, CP, MALFUNCTION
→
REPAIR
→
RECALIBRATE
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE
For Life Sciences
Only
STATUS..
SERVICE INFORMATION:
222
PRESSURE IN CHAMBER
→
2 PSIG PRESSURE SENSED IN CHAMBER
WHEN NOT IN CYCLE
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
REPAIR S02
2.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3.
TRANSDUCER, CP, MALFUNCTION
→
REPAIR
→
RECALIBRATE
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE CONTROL BOARD*
→
RECALIBRATE
PAPER
EXIT
FEED
For Healthcare
Only
*Depending upon the configura-
tion of the unit, screen text
may reference 146656-185
control board, or 146659-065
control board.
Screen with Service Instructions