5-14
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Service Instructions
Chamber Pressure/
Temperature Failure,
Continued
STATUS..
SERVICE INFO: CHAMBER
220
PRESSURE/TEMPERATURE FAILURE
→
PRESSURE OR TEMPERATURE OUTSIDE
NORMAL STEAM RANGE
CAUSES AND CORRECTION:
1.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
2.
TRANSDUCER, CP, MALFUNCTION
→
REPAIR
3.
RTD PROBE, CT, MALFUNCTION
→
REPAIR
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE CONTROL BOARD*
→
RECALIBRATE
PAPER
EXIT
FEED
For Healthcare
Only
*Depending upon the configura-
tion of the unit, screen text
may reference 146656-185
control board, or 146659-065
control board.
For Life Sciences
Only
STATUS..
SERVICE INFO: CHAMBER
220
PRESSURE/TEMPERATURE FAILURE
→
PRESSURE OR TEMPERATURE OUTSIDE
NORMAL STEAM RANGE
CAUSES AND CORRECTION:
1.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
2.
TRANSDUCER, CP, MALFUNCTION
→
REPAIR
3.
RTD PROBE, CT, MALFUNCTION
→
REPAIR
4.
MAIN CONTROL FAILURE
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
REPLACE 146656-185 CONTROL BOARD
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE