5-12
129367-410
Routine Maintenance
Troubleshooting
Alarm
Description
Screen with Operator Instructions
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
TRY TO EXHAUST CHAMBER ACCORDING
TO OPTIMAL COOLING RATE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
Exhaust Rate Too Fast
Occurs if chamber
pressure is 1.0 psig
above the fixed rate
during slow exhaust
phase. (Sterilizers with
Liquid cycle option
only.)
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
241
EXHAUST RATE TOO FAST
Screen with Service Instructions
Screen with Operator Instructions
Door Unsealed
Occurs if steam pres-
sure in door seal drops
below 10 psig.
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
207
DOOR UNSEALED
CHAMBER:
000.0 C
00.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
•
EXHAUST OR AIR BREAK CHAMBER TO
ATMOSPHERIC PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS..
SERVICE INFORMATION:
242
EXHAUST RATE TOO FAST
→
CHAMBER EXHAUSTED FASTER THAN
THAN THE EXPECTED RATE
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUCNTION
→
REPAIR S40
2.
SOLENOID VALVE MALFUCNTION
→
REPAIR S03
3.
CHAMBER STEAM TRAP MALFUNCTION
→
REPAIR
4.
CONTROL OUT OF CALIBRATION
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE