ROUTINE MAINTENANCE
Amsco
®
Century
Series
Sterilizers
(05/27/05)
P129367-410
Printed in U.S.A.
© 2005, STERIS Corporation, All rights reserved.
Страница 1: ...Table of Contents RoutineMaintenance 129367 410 ROUTINE MAINTENANCE Amsco Century Series Sterilizers 05 27 05 P129367 410 Printed in U S A 2005 STERIS Corporation All rights reserved ...
Страница 2: ...ety Precautions to be observed when operating or servicingthissterilizercanbefoundinSection1ofthismanual Donotoperate or service the sterilizer until you have become familiar with this information Thissterilizerisnotdesignedtoprocessflammablecompounds Anyalteration of the sterilizer which affects its operation will void the warranty could adversely affect sterilization efficacy could violate feder...
Страница 3: ...ner 3 6 3 2 3CleanChamber 3 6 3 3 WeeklyMaintenance 3 7 3 4 AsRequired 3 8 3 4 1 Change Printer Paper Roll 3 8 3 4 2 Change Printer Ink Cartridge 3 10 3 4 3 Replace Door Seal 3 12 4 SERVICEPROCEDURES 4 1 4 1 Air Filter Replacement 4 1 4 2 CleanStrainers 4 1 4 3 SteamTrapReplacement 4 3 4 4 Valves 4 3 4 4 1 Clean or Replace Piping Check Valves 4 3 4 4 2 Clean or Replace Manifold Check Valves 4 3 4 ...
Страница 4: ...g before beginning work on the steam trap Proper testing of the safety valve requires the valve to be under pressure Exhaust from the safety valve is hot and can cause burns Proper safety attire gloves eye protection insulated overalls as designated by OSHA are required Testing is to be performed by qualified service personnel only Sterilizeroperatormaybeseverelyburnedbyscaldingwaterifthewaterleve...
Страница 5: ... door CAUTION POSSIBLE EQUIPMENT DAMAGE Insufficient service clearance will make repairs more difficult and time consuming Piping sized too small may cause water hammer resulting in damage to the sterilizer Neveruseawirebrush abrasivesorsteelwoolondoorandchamberassembly Donotusecleanerscontaining chloride on stainless steel surfaces Chloride based cleaners will deteriorate stainless steel eventual...
Страница 6: ...out capability Backflow prevention is by others Water Pressure measured specification is 30 to 50 psig static and dynamic Waterpressuresuppliedmustbewithinspecificationsasshown on the Equipment Drawing If pressure is too high a regulator must be installed If water pressure is too low equipment performance will be affected use of a booster pump by others is recommended Water Quality supplied must b...
Страница 7: ...2 3 Electrical Service Door runs up and down smoothly Door open close time is 7 0 9 0 seconds power door units Door up switches adjusted correctly Chamber strainer in place Rack and shelves loading car operates correctly Paper loaded in printer Unit powers up correctly If prevacuum sterilizer run leak test cycle leak rate is to be less than 1 0 mmHg minute Verify operation of a typical cycle e g G...
Страница 8: ...Steam Supply Valve Main Power Disconnect Century Sterilizer GeneratorWater Supply Valve Sight Glass GeneratorDrainValve Century Control Panel CenturyControl TouchScreen Century Control with Door Open Printer Optional Electric Steam Generator Figure 2 1 Typical Amsco Century Series Sterilizer ...
Страница 9: ...STEAM AND WATER SUPPLY HAND VALVE 3 1 Inspect each hand valve for smooth operation and proper valve seating X X X X X X 3 2 Inspect valve for leaks X X X X X X MaintenanceproceduresdescribedinSections3and4shouldbeperformed regularly at the intervals indicated using the maintenance schedule in Table 3 1asaguide Localconditions waterquality usage etc mayrequire more frequent maintenance than indicat...
Страница 10: ...k chamber and jacket pressure gauges for accuracy X 5 7 Replace air inlet filter cartridge X 5 8 Water Valves rebuild valve s X 6 0 CONTROL 6 1 Check cooling fan operation X X X X X X 6 2 Replace fan filter X 6 3 Confirmcontrolcalibration X 6 4 Replacebattery A R 6 5 Verify operation of flood alarm control X X X X X X 6 6 Replaceprinter A R 6 7 Replaceribboncartidge 7 0 ELECTRIC STEAM GENERATOR IF...
Страница 11: ... X X X X X X 7 22 Clean heating element and boiler chamber X X X X X X 7 23Descalegenerator X X X 7 24 Determine water hardness compare to specification X X 7 25 Test safety valve for proper operation X ATTENTION Warranty on this steam generator will be voided unless daily flushing procedures are performed See Daily Maintenance Instructions in this manual for proper instruction or call STERIS for ...
Страница 12: ...e flushing 7 2 FLUSHGENERATOR STOP TIMER START TIMER CANCEL THE GENERATOR SHOULD BE FLUSHED EVERY DAY TO FLUSH GENERATOR 1 VERIFY GENERATOR PRESSURE IS 0 psig 2 OPEN GENERATOR MANUAL DRAIN VALVE 3 PRESS START TIMER TO BEGIN FLUSH PHASE 4 ALLOW GENERATOR TO FLUSH FOR 5 MINUTES FLUSH TIMER 00 00 Pressing STOP TIMER stops flush phase and resets timer to three minutes Pressing CANCEL bypasses flush ph...
Страница 13: ...screen 73 Generatorautomaticallyfillstotheproper level and starts to heat Display screen advances to the Main Menu screen 1 Allow 10 minutes warm up time once generator starts to fill 10 Close front cabinet panel NOTE The generator must be flushed every day before use 73 FLUSHING OF GENERATOR COMPLETE TO RESTART GENERATOR 1 CLOSE GENERATOR MANUAL DRAIN VALVE 2 PRESS CONTINUE TO START THE GENERATOR...
Страница 14: ...ed down and rinsed following any spills or other soiling 1 If applicable the shelf assembly must be removed before cleaning the chamber Single Door a Use a 1 1 8 inch socket wrench to loosen and remove the rear retaining nut see Fig 3 3 on the shelf assembly b Use a 1 8 inch hex wrench to loosen but don t remove the six set screws see Fig 3 3 at the front of the shelf assembly c Carefully slide th...
Страница 15: ...Criti Klenz Instrument Detergent or current STERIS equivalent into drain Solution may puddle in the bottom of the chamber d Wait five minutes e Close the door turn the steam valve on and run a Gravity cycle set for one minute sterilize and one minute dry f Open door and place strainer back in drain Figure 3 3 Remove Chamber Shelf Assembly CAUTION POSSIBLE EQUIPMENT DAMAGE Never use a wire brush ab...
Страница 16: ... remove used paper roll from spindle 4 Open access door and remove old paper roll gently pulling any remaining tape up and out of printer Access Door 1 Tear paper between take up spool and printer The printer paper roll should be changed whenever a colored stripe is visible on one or both edges of the printout paper 3 4 As Required 3 4 1 Change Printer Paper Roll Remove Used Tape Printer Behind In...
Страница 17: ... STATUS ABORT 7 Press PAPER FEED touch screen pad on display until paper advances through printer and ink cartridge exiting the front 8 Continue pressing PAPER FEED or pull tape gently until about 18 457 mm of paper hangs out of printer Insert end of paper into slot of take up spool then replace right end of spool Pull Tape Gently Paper Advances Through Printer ...
Страница 18: ...idler Manually roll up slack paper 1 Tear paper between take up spool and printer 2 Open access door then press on right end of ink cartridge until left end of cartridge pops out of the printer The printer ink cartridge should be changed as soon as the type on printouts is light or faded and before printouts become difficult to read 3 4 2 Change Printer Ink Cartridge ...
Страница 19: ...g sure paper tape slides between ink cartridge housing and ink ribbon 5 Retighten ribbon by rotating wheel on left side of cartridge 1 4 turn Then see Section 3 4 1 Change Printer Paper Roll Steps 8 through 10 to reinstall take up spool 4 Install left end of cartridge first then push right end in as shown snapping it into place Rotate Wheel 1 4 turn to tighten ribbon ...
Страница 20: ...he support stand align top and bottom indicators with the ports in the top and bottom of the end ring frame b Press seal in at each reference point with fingertips c Press remainder of the seal into end ring groove starting at the corners 5 Test installation a Attempt to close the door If the door jams or sticks at any point in its travel check to ensure the seal has been fully pressed into the gr...
Страница 21: ...Depressor to Pry Out Door Seal Reference Indicator Enlarged Figure 3 4b Location of Seal Date Information and Reference Indicators Location of Reference Indicators Location of Seal Date Information Figure 3 4a Remove Door Seal Rear of Door Seal Location of Reference Indica tors ...
Страница 22: ...em is exposedtocontaminationwheneverthefilterortheairlinesbelowthefilterare opened Keep these components as clean as possible when servicing The bacterial air filter contains a replaceable filter cartridge refer to Section 3 1 PeriodicMaintenanceScheduleforfrequency 1 Remove the bowl from the filter body by loosening the retaining nut 2 Remove the old filter element and discard 3 Insert the new fi...
Страница 23: ...4 2 129367 410 RoutineMaintenance ServiceProcedures Figure 4 1 Serviceable Components ...
Страница 24: ...n firmly Use a socket head wrench to tighten NOTE Seat and diaphragm are a lapped pair 2 Install new diaphragm 3 Replacecapandattacheddiaphragmassembly usinganewgasket 4 Check for leaks 4 3 Steam Trap Replacement 4 4 1 Clean or Replace Piping Check Valves 4 4 2 Clean or Replace Manifold Check Valves Repair of check valves is limited to cleaning of valve seats and or replacing seal rings when forei...
Страница 25: ...ver to snap shut 4 4 4 Safety Valve Test WARNING BURN HAZ ARD Propertestingofthe safety valve requires the valve to be under pres sure Exhaust from the safetyvalveishotandcan causeburns Propersafety attire gloves eyeprotec tion insulatedoveralls as designated by OSHA is required Testing is to be performed by qualified service personnel only CAUTION POSSIBLE EQUIPMENT DAMAGE Actuationatlessthan75 o...
Страница 26: ...ELECTRIC SHOCK HAZARD Disconnect all utilities to sterilizer before servicing See Section 1 for additional warnings WARNING BURN HAZ ARD Beforeperformingany cleaning or maintenance procedures allow sterilizer andgenerator ifapplicable tocooltoroomtemperature 5 1 1 Typical Alarm Screen STATUS PAPER PRINT FEED SILENCE SERVICE ALARM HELP CHAMBER 000 0 C 00 0 psig STERILIZER WILL AUTOMATICALLY TRY TO ...
Страница 27: ...PER FEED advances the printer paper up by one line PressingSERVICEHELPadvancesdisplaytothecorrespondingservice informationscreen Thisscreenprovidesthequalifiedservicetechnician with possible causes and advanced corrective actions for that alarm condition IMPORTANT In the event of an alarm condition the operator should always follow the instructions indicated on the alarm screen When an alarm occur...
Страница 28: ...WITHIN ALLOTTED TIME CAUSES AND CORRECTION 1 STEAM PRESSURE LESS THAN 50 PSIG CHECK STEAM SUPPLY PIPING 2 STEAM REGULATOR MALFUNCTION REPAIR 3 SOLENOID VALVE MALFUNCTION REPAIR S09 REPAIR S02 4 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON STATUS SERVICE INFORMATION 206 TOO LONG IN CHARGE ABORT PAPER EXIT FEED For Healthcare Only CHAMBER DID NOT REACH STERILIZE TEMPERATUR...
Страница 29: ...MOSPHERIC PRESSURE WITHIN ALLOTTED TIME CAUSES AND CORRECTION 1 CHAMBER DRAIN STRAINER PLUGGED CLEAN 2 SOLENOID VALVE MALFUNCTION REPAIR S03 3 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON STATUS SERVICE INFORMATION 201 TOO LONG IN EXHAUST PAPER SERVICE EXIT FEED MODE For Life Sciences Only CHAMBER DID NOT EXHAUST TO ATMOSPHERIC PRESSURE WITHIN ALLOTTED TIME CAUSES AND CO...
Страница 30: ... 0 psig STERILIZER WILL AUTOMATICALLY TRY TO COMPLETE CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CHECK WATER SUPPLY VALVE IF CLOSED OPEN VALVE 3 IF ALARM RECURS ABORT CYCLE AND CALL SERVICE STATUS ALARM 202 TOO LONG IN EVACUATION SILENCE ABORT SERVICE ALARM HELP ABORT STATUS PAPER MORE PRINT FEED HELP STATUS SERVICE INFORMATION 203 TOO LONG IN EVACUATION CHAMBER DID NOT REACH REQUIRED VACUUM LE...
Страница 31: ...EAL NOT ACTIVATED CHECK SEAL CHECK SEAL STEAM AND EXHAUST For Healthcare Only CHAMBER DID NOT REACH REQUIRED VACUUM LEVEL WITHIN ALLOTTED TIME CAUSES AND CORRECTION 6 LEAK IN PLUMBING REPAIR RUN A LEAK TEST 7 CONTROL OUT OF CALIBRATION RECALIBRATE CONTROL CONTACT QUALIFIED SERVICE PERSON STATUS SERVICE INFORMATION 204 TOO LONG IN EVACUATION PAPER SERVICE EXIT FEED MODE For Life Sciences Only CHAMB...
Страница 32: ...E INFORMATION 226 TOO LONG IN AIR BREAK CHAMBER DID NOT AIR BREAK VACUUM TO 2 inHg WITHIN ALLOTTED TIME CAUSES AND CORRECTION 1 AIR INLET FILTER PLUGGED REPLACE 2 SOLENOID VALVE MALFUNCTION REPAIR S01 3 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON PAPER SERVICE EXIT FEED MODE For Life Sciences Only For Healthcare Only STATUS SERVICE INFORMATION 226 TOO LONG IN AIR BREAK ...
Страница 33: ...zer will auto matically resume sterilize time after set temperature is reached For Life Sciences Only Occurs if chamber temperature drops below sterilize temperature Under Sterilize Temperature STATUS PAPER PRINT FEED SILENCE SERVICE ALARM HELP STATUS ALARM 223 UNDER STERILIZE TEMPERATURE CHAMBER 000 0 C 00 0 psig STERILIZER WILL AUTOMATICALLY RESTART STERILIZE TIME AND CONTINUE CYCLE AFTER SET TE...
Страница 34: ...ilizer will auto matically resume sterilize time after set temperature is reached For Healthcare Only For Life Sciences Only Occurs if chamber temperature exceeds the maximum sterilize temperature control temperature plus over temperature value SILENCE SERVICE ALARM HELP CHAMBER 000 0 C 00 0 psig STERILIZER WILL AUTOMATICALLY CONTINUE CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 IF ALARM RECURS A...
Страница 35: ...MPERATURE IS ABOVE SETPOINT BY MORE THAN PRESCRIBED AMOUNT CAUSES AND CORRECTION 1 STEAM PRESSURE MORE THAN 50 PSIG CHECK STEAM SUPPLY PIPING 2 CHAMBER STEAM TRAP MALFUNCTION REPAIR 3 SOLENOID VALVE MALFUNCTION REPAIR S09 REPAIR S02 4 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON SERVICE STATUS PAPER EXIT MODE PRINT FEED For Life Sciences Only STATUS SERVICE INFORMATION 2...
Страница 36: ...h Service Instructions STATUS SERVICE INFORMATION 234 TOO LONG IN STERILIZE THE TIME SPENT IN THE STERILIZE PHASE WAS TOO LONG CAUSES AND CORRECTION 1 SOLENOID VALVE MALFUNCTION REPAIR S02 2 SOLENOID VALVE MALFUNCTION REPAIR S09 3 SOLENOID VALVE MALFUNCTION REPAIR S40 4 CHAMBER STEAM TRAP MALFUNCTION REPAIR 5 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON PAPER SERVICE EXI...
Страница 37: ...T Screen with Service Instructions Screen with Operator Instructions Door Unsealed Occurs if steam pres sure in door seal drops below 10 psig STATUS PAPER PRINT FEED SILENCE SERVICE ALARM HELP STATUS ALARM 207 DOOR UNSEALED CHAMBER 000 0 C 00 0 psig STERILIZER WILL AUTOMATICALLY ABORT CYCLE EXHAUST OR AIR BREAK CHAMBER TO ATMOSPHERIC PRESSURE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE ST...
Страница 38: ...SSURE IN DOOR SEAL BELOW 10 PSIG CAUSES AND CORRECTION 1 SEAL PRESSURE SWITCH MALFUNCTION CHECK PS1 PS2 CONNECTIONS READJUST PS1 PS2 REPAIR PS1 PS2 2 SEAL NOT ACTIVATED CHECK SEAL STEAM CHECK SEAL EXHAUST 3 SOLENOID VALVE MALFUNCTION REPAIR S35 S36 STATUS SERVICE INFORMATION 208 DOOR UNSEALED PAPER CLEAR EXIT FEED ALARM Door Unsealed Continued For Life Sciences Only STEAM PRESSURE IN DOOR SEAL BEL...
Страница 39: ...NTROL BOARD POWER STATUS LEDs REPLACE CONTROL BOARD RECALIBRATE PAPER EXIT FEED For Healthcare Only Depending upon the configura tion of the unit screen text may reference 146656 185 control board or 146659 065 control board For Life Sciences Only STATUS SERVICE INFO CHAMBER 220 PRESSURE TEMPERATURE FAILURE PRESSURE OR TEMPERATURE OUTSIDE NORMAL STEAM RANGE CAUSES AND CORRECTION 1 CONTROL OUT OF C...
Страница 40: ...N 1 C CAUSES AND CORRECTION 1 LOOSE CONNECTION IN PROBE WIRING REPAIR 2 CHAMBER REFERENCE PROBE FAILED REPLACE RECALIBRATE 3 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON 4 MAIN CONTROL FAILURE CHECK CONTROL BOARD POWER STATUS LEDs REPLACE 146656 185 CONTROL BOARD REPLACE 146659 003 ANALOG BOARD RECALIBRATE STATUS SERVICE INFORMATION 250 REFERENCE CHAMBER DEVIATION PAPER ...
Страница 41: ...ROL BOARD POWER STATUS LEDs REPLACE 146656 185 CONTROL BOARD RECALIBRATE PAPER SERVICE EXIT FEED MODE For Life Sciences Only STATUS SERVICE INFORMATION 216 WASTE TEMP PROBE FAILURE RTD PROBE WT OUTPUT IS OUTSIDE NORMAL RANGE CAUSES AND CORRECTION 1 LOOSE CONNECTION IN PROBE WIRING REPAIR 2 PROBE FAILED REPLACE RECALIBRATE 3 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON 4 ...
Страница 42: ...ERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CHECK DOOR FOR OBSTRUCTION REMOVE OBSTRUCTION AND CLOSE DOOR 3 IF DOOR WILL NOT CLOSE CALL SERVICE Screen with Service Instructions STATUS SERVICE INFORMATION 240 TOO LONG TO CLOSE DOOR DOOR SWITCH DID NOT MAKE IN ALLOTTEDTIME CAUSES AND CORRECTION 1 DOOR SWITCH MALFUNCTION CHECK LS1 LS2 CONNECTIONS READJUST LS1 LS2 REPAIR LS1 LS2 2 POWER DOOR MECHANISM FAILUR...
Страница 43: ...on Screen with Operator Instructions STATUS SERVICE INFORMATION 246 TOO LONG TO OPEN DOOR PAPER CLEAR EXIT FEED ALARM DOOR SWITCH DID NOT OPEN IN ALLOTTEDTIME CAUSES AND CORRECTION 1 DOOR SWITCH MALFUNCTION CHECK LS1 LS2 CONNECTIONS READJUST LS1 LS2 2 POSER DOOR MECHANISM FAILURE REPAIR MECHANISM REPLACE MOTOR For Life Sciences Only STATUS SERVICE INFORMATION 246 TOO LONG TO OPEN DOOR PAPER EXIT F...
Страница 44: ... RECALIBRATE CONTACT QUALIFIED SERVICE PERSON 3 TRANSDUCER CP MALFUNCTION REPAIR RECALIBRATE 4 MAIN CONTROL FAILURE CHECK CONTROL BOARD POWER STATUS LEDs REPLACE 146656 185 CONTROL BOARD RECALIBRATE PAPER SERVICE EXIT FEED MODE For Life Sciences Only STATUS SERVICE INFORMATION 222 PRESSURE IN CHAMBER 2 PSIG PRESSURE SENSED IN CHAMBER WHEN NOT IN CYCLE CAUSES AND CORRECTION 1 SOLENOID VALVE MALFUNC...
Страница 45: ...OOSE CONNECTION IN PROBE WIRING REPAIR 2 PROBE FAILED REPLACE RECALIBRATE 3 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON 4 MAIN CONTROL FAILURE CHECK CONTROL BOARD POWER STATUS LEDs REPLACE 146656 185 CONTROL BOARD RECALIBRATE PAPER SERVICE EXIT FEED MODE For Life Sciences Only STATUS SERVICE INFORMATION 216 WASTE TEMP PROBE FAILURE RTD PROBE WT OUTPUT IS OUTSIDE NORMAL ...
Страница 46: ... Service Instructions STATUS SERVICE INFORMATION 210 WATER IN CHAMBER EXCESS WATER SENSED IN CHAMBER WARNING BURN HAZARD CAUSES AND CORRECTION 1 JACKET STRAINER PLUGGED CLEAN 2 JACKET TRAP FAILED CLOSED REPAIR 3 CHAMBER TRAP FAILED CLOSED REPAIR 4 WATER ENTERED CHAMBER THROUGH STEAM PIPING CHECK BOILER OR STEAM GENERATOR REPAIR PIPING 5 WATER FLOAT SENSOR MALFUNCTION REPAIR PAPER SERVICE EXIT FEED...
Страница 47: ...ARGE JACKET DID NOT REACH REQUIRED TEMPERATURE WITHIN ALLOTTED TIME CAUSES AND CORRECTION 1 STEAM PRESSURE LESS THAN 50 PSIG CHECK STEAM SUPPLY PIPING 2 STEAM REGULATOR MALFUNCTION REPAIR 3 SOLENOID VALVE MALFUNCTION REPAIR S09 4 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON PAPER SERVICE EXIT FEED MODE For Life Sciences Only STATUS SERVICE INFORMATION 232 TOO LONG IN JAC...
Страница 48: ... CALL SERVICE STATUS SERVICE INFORMATION 228 TOO LONG TO SEAL DOOR DOOR SEAL DID NOT REACH 10 PSIG WITHIN ALLOTTED TIME CAUSES AND CORRECTION 1 SEAL PRESSURE SWITCH MALFUNCTION READJUST PS1 PS2 REPAIR PS1 PS2 2 SEAL NOT ACTIVATING CHECK SEAL STEAM CHECK SEAL EXHAUST 3 SOLENOID VALVE MALFUNCTION REPAIR S35 S36 PAPER CLEAR EXIT FEED ALARM For Life Sciences Only For Life Sciences Only STATUS PAPER PR...
Страница 49: ... Unseal Door STATUS PAPER PRINT FEED SILENCE SERVICE ALARM HELP STATUS ALARM 229 TOO LONG TO UNSEAL DOOR CHAMBER 000 0 C 00 0 psig STERILIZER WILL REMAIN IN ALARM CONDITION UNTIL DOOR IS UNSEALED OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE 3 IF LOAD MUST BE REMOVED REFER TO EMERGENCY DOOR OPERATION PROCEDURE IN OPERATING MANUAL STATUS SERVICE INFORMATION 230 TOO LONG TO UNSEAL DOOR DOOR S...
Страница 50: ...NSDUCER FAILURE CHAMBER 000 0 C 00 0 psig STERILIZER WILL AUTOMATICALLY ABORT CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE STATUS SERVICE INFORMATION 230 TOO LONG TO UNSEAL DOOR DOOR SEAL PRESSURE NOT BELOW 10 PSIG WITHIN ALLOTTED TIME CAUSES AND CORRECTION 1 SEAL PRESSURE SWITCH MALFUNCTION READJUST PS1 PS2 REPAIR PS1 PS2 2 SEAL NOT RETRACTING CHECK SEAL STEAM CHECK SEAL EXHAUST 3 S...
Страница 51: ...outside the normal range Chamber Temperature Probe Failure STATUS PAPER PRINT FEED SILENCE SERVICE ALARM HELP STATUS ALARM CHAMBER 211 TEMPERATURE PROBE FAILURE CHAMBER 000 0 C 00 0 psig STERILIZER WILL AUTOMATICALLY ABORT CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE For Life Sciences Only STATUS SERVICE INFO CHAMBER 218 PRESSURE TRANSDUCER FAILURE TRANSDUCER CP OUTPUT VOLTAGE IS OUT...
Страница 52: ...ROL BOARD POWER STATUS LEDs REPLACE CONTROL BOARD RECALIBRATE PAPER EXIT FEED For Healthcare Only Depending upon the configura tion of the unit screen text may reference 146656 185 control board or 146659 065 control board STATUS SERVICE INFORMATION 212 CHAMBER TEMP PROBE FAILURE RTD PROBE CT OUTPUT IS OUTSIDE NORMAL RANGE CAUSES AND CORRECTION 1 LOOSE CONNECTION IN PROBE WIRING REPAIR 2 PROBE FAI...
Страница 53: ...N PROBE WIRING REPAIR 2 PROBE FAILED REPLACE RECALIBRATE 3 CONTROL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED SERVICE PERSON 4 MAIN CONTROL FAILURE CHECK CONTROL BOARD POWER STATUS LEDs REPLACE 146656 185 CONTROL BOARD RECALIBRATE PAPER SERVICE EXIT FEED MODE For Life Sciences Only STATUS SERVICE INFORMATION 214 JACKET TEMP PROBE FAILURE RTD PROBE JT OUTPUT IS OUTSIDE NORMAL RANGE CAUSES AND...
Страница 54: ...NS 1 SILENCE ALARM 2 CALL SERVICE Screen with Service Instructions STATUS SERVICE INFORMATION 238 DOOR SWITCH FAILURE DOOR SWITCH OPEN WHILE SEAL SWITCH CLOSED CAUSES AND CORRECTION 1 DOOR SWITCH MALFUNCTION CHECK LS1 LS2 CONNECTIONS READJUST LS1 LS2 REPAIR LS1 LS2 2 SEAL PRSSURE SWITCH MALFUNCTION READJUST PS1 PS2 REPAIR PS1 PS2 PAPER SERVICE EXIT FEED MODE For Life Sciences Only For Healthcare O...
Страница 55: ...sig LOAD 000 0 C STERILIZER WILL AUTOMATICALLY ABORT CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE Screen with Service Instructions STATUS SERVICE INFORMATION 248 LOAD TEMP PROBE FAILURE RTD PROBE LT OUTPUT IS OUTSIDE NORMAL RANGE CAUSES AND CORRECTION 1 LOOSE CONNECTION IN PROBE WIRING REPAIR 2 PROBE FAILED REPLACE RECALIBRATE 3 CONTOL OUT OF CALIBRATION RECALIBRATE CONTACT QUALIFIED...
Страница 56: ...CORRECTION 1 LOOSE CONNECTION BETWEEN BOARD 1 2 3 AND CONNECTION BOARD RESEAT BOARD 2 LOOSE CONNECTION BETWEEN 146659 065 BOARD AND CONNECTOR BOARD RESEAT BOARD 3 BOARD 1 2 OR 3 DIP SWITCHES CONFIGURED INCORRECTLY REFER TO THE SERVICE MANUAL FOR INFORMATION ON DIP SWITCH SETTINGS OR CALL SERVICE 4 BOARD 1 2 3 OR 146659 065 FAIL REPLACE 5 FAULTY CONNECTION IN CONTROL BOX REPLACE 146657 782 PAPER EX...
Страница 57: ... FEED SILENCE ALARM STATUS ALARM LOAD 278 RAM FAILURE CHAMBER 300 0 F 0 0 psig STERILIZER WILL AUTOMATICALLY ABORT CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE RAM Failure Occurs if random access memory on the main control board fails STATUS PAPER PRINT FEED SILENCE ALARM STATUS ALARM LOAD 279 ADC FAILURE CHAMBER 300 0 F 0 0 psig STERILIZER WILL AUTOMATICALLY ABORT CYCLE OPERATOR INS...
Страница 58: ...OAD 280 BOARD OVER TEMP FAILURE CHAMBER 300 0 F 0 0 psig STERILIZER WILL AUTOMATICALLY ABORT CYCLE OPERATOR INSTRUCTIONS 1 SILENCE ALARM 2 CALL SERVICE Board Over Temp Failure Occurs if temperature of the main control board exceeds acceptable conditions Alarm Description Screen with Operator Instructions STATUS SERVICE INFORMATION 287 DOOR A SEAL SWITCH MALFUNCTION SEAL SWITCH CLOSED OUT OF CYCLE ...