IEEE-488 Common Commands
Bit
Number
Bit Name
Decimal
Value
Definition
4
Execution
Error
16
An execution error occurred.
5
Command
32
A command syntax error occurred.
6
(not used)
64
(Reserved for future use)
7
Power On
128
Power has been cycled since the last time the event register was read or
cleared.
Status Byte Register
The following table describes the Status Byte Register.
Bit
Number
Bit Name
Decimal
Value
Definition
0
(not used)
1
(Reserved for future use)
1
(not used)
2
(Reserved for future use)
2
Error Queue
4
One or more errors in the Error Queue. Use
to read and
delete errors.
3
Questionable
Data Summary
8
One or more bits are set in the Questionable Data Register (bits must be
enabled, see
).
4
Message Avail-
able
16
Data is available in the instrument's output buffer.
5
Standard
Event Sum-
mary
32
One or more bits are set in the Standard Event Register (bits must be ena-
bled, see
).
6
Master Sum-
mary
64
One or more bits are set in the Status Byte Register and may generate a
Request for Service (RQS). Bits must be enabled using
.
7
Operation Reg-
ister
128
One or more bits are set in the Operation Status Register. Bits are ena-
bled using
.
*CLS
Clear Status Command
. Clears the event registers in all register groups. Also clears the error queue.
Parameter
Typical Return
(none)
(none)
Clear event register bits and error queue:
*CLS
230
Agilent 33500 Series Operating and Service Guide