Output Configuration
l
Front Panel:
Then choose one of the following options. The
Symmetry
option allows you to use the knob or keypad to specify a
value.
If you use the keypad, press
Percent
to finish.
l
SCPI:
FUNCtion:RAMP:SYMMetry {<percent>|MIN|MAX}
The
command sets the symmetry to 100%.
Voltage Autoranging
Autoranging is enabled by default and the instrument selects optimal attenuator settings. With autoranging disabled,
the instrument uses the current attenuator settings and does not switch attenuator relays.
l
You can disable autoranging to eliminate momentary disruptions caused by attenuator switching while changing
amplitude. However:
l
The amplitude and offset accuracy and resolution (and waveform fidelity) may be adversely affected when reduc-
ing the amplitude below a range change that would occur with autoranging on.
l
You may not achieve minimum amplitude with autoranging on.
l
Some instrument specifications do not apply with autoranging off.
l
Front Panel:
Agilent 33500 Series Operating and Service Guide
63