FUNCtion Subsystem
l
Limits due to Output Coupling: If two channels are coupled, limitations of setting the levels of both channels will be
checked before a change in level is executed. In this case, if a change in level would exceed a LIMIT setting, or
instrument output specifications for either channel, the level is clipped to the maximum (or minimum) allowable
value and a "Settings conflict" error will be generated.
l
To invert the waveform relative to the offset voltage, use
.
[SOURce[1|2]:]FUNCtion:ARBitrary:SRATe {<
sample_rate
>|MINimum|MAXimum}
[SOURce[1|2]:]FUNCtion:ARBitrary:SRATe? {MINimum|MAXimum}
Sets the sample rate for the arbitrary waveform.
Parameter
Typical Return
1 µSa/s to 250 MSa/s, default 40 kSa/s. Limited to 62.5 MSa/s if FUNC-
tion:ARBitrary:FILTer is OFF.
+1.00000000004
Set sample rate to 10 kSa/s:
FUNC:ARB:SRAT 1e4
l
The sample rate and frequency parameter are not coupled when playing an arbitrary waveform segment. The con-
cept of frequency does not apply for arbitrary waveform sequences.
l
Setting a sample rate when not in the ARB mode will not change the frequency. For example, if the current func-
tion is sine, setting sample rate has no effect until the function changes to ARB.
l
The maximum sample rate depends on the filter applied to the arbitrary waveform. See
for details.
[SOURce]:FUNCtion:ARBitrary:SYNCh
Causes two independent arbitrary waveforms to synchronize to first point of each waveform (two-channel instru-
ments only).
Parameter
Typical Return
(none)
(none)
Load an internal haversine waveform into channel 1 and a custom arbitrary waveform from a USB drive into channel
2. Set both sample rates to 100 kSa/s and then synchronizes both channels to the first point of each waveform:
MMEM:LOAD:DATA "Int:\Builtin\HAVERSINE.arb"
FUNC:ARB "Int:\Builtin\HAVERSINE.ARB"
FUNC ARB
FUNC:ARB:SRATE 1E+05
MMEM:LOAD:DATA2 "USB:\MyFiles\TestDUT3.arb"
SOUR2:FUNC:ARB "USB:\MyFiles\TestDUT3.arb"
SOUR2:FUNC ARB
SOUR2:FUNC:ARB:SRAT 1E+05
FUNC:ARB:SYNC
216
Agilent 33500 Series Operating and Service Guide
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......