Frequency Sweep
l
Front Panel:
l
SCPI:
[SOURce[1|2]:]FREQuency:STARt {<
[SOURce[1|2]:]FREQuency:STOP {<
Center Frequency and Frequency Span
You can also set the sweep frequency boundaries of the sweep using a center frequency and frequency span. These
parameters are similar to the start frequency and stop frequency (above) and they provide added flexibility.
l
Center frequency: 1 µHz to 30 MHz (limited to 200 kHz for ramps). The default is 550 Hz.
l
Frequency span: -30 µHz to 30 MHz (limited to 200 kHz for ramps). The default is 900 Hz.
l
To sweep up in frequency, set a positive frequency span; to sweep down, set a negative frequency span.
l
For sweeps with Marker Off, the
Sync
signal is a square waveform with a 50% duty cycle. (missing or bad snippet)
The frequency of the
Sync
waveform is equal to the specified sweep time. The signal is output from the front-
panel
Sync
connector.
l
For frequency sweeps with Marker On, the
Sync
signal is a TTL "high" at the beginning of the sweep and a "low" at
the marker frequency. You can change this with
.
l
Front Panel:
92
Agilent 33500 Series Operating and Service Guide
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......