Pulse Waveforms
Enter the Duty Cycle:
l
SCPI:
Edge Times
The edge times set the transition times for the leading and trailing edges of the pulse, either independently or together.
The edge time represents the time between the 10% and 90% thresholds.
l
Edge time: 8.4 ns to 1 μs (default 10 ns).
l
The specified edge time must fit within the specified pulse width
. The instrument will adjust the
edge time to accommodate the specified pulse width.
l
Front Panel:
l
SCPI:
FUNC:PULS:TRAN:LEAD {<seconds>|MIN|MAX}
FUNC:PULS:TRAN:TRA {<seconds>|MIN|MAX}
FUNC:PULS:TRAN[:BOTH] {<seconds>|MIN|MAX}