Front-Panel Menu Operation Introduction
1.
Select the function and amplitude for the burst.
For burst waveforms, you can select sine, square, ramp, pulse, arbitrary waveforms, triangle, or PRBS. Noise is
allowed only in the "gated" burst mode and DC is not allowed. For this example, select a sine wave with an ampli-
tude of 5 Vpp.
2.
Select the burst mode.
Press
Burst
and then press the
Burst Off / On
softkey. Notice that a status message
N Cycle Burst, Trig Imm
is shown in the tab of the current channel.
3.
Set the burst count.
Press the
# of Cycles
softkey and then set the count to "3" using the numeric keypad or knob. Press the
Enter
34
Agilent 33500 Series Operating and Service Guide