Amplitude Modulation (AM) and Frequency Modulation (FM)
Carrier Waveform Shape
l
AM or FM carrier shape: Sine (default), Square, Ramp, Pulse, Triangle, Noise, PRBS, or Arbitrary waveform. You can-
not use DC as the carrier waveform.
l
For FM, the carrier frequency must always be greater than or equal to the frequency deviation. Attempting to set a
deviation greater than the carrier frequency will cause the instrument to set the deviation equal to the carrier
frequency.
l
The carrier frequency plus the deviation cannot exceed the selected function's maximum frequency plus 100 kHz.
If you attempt to set the deviation to an invalid value, the instrument adjusts it to the maximum value allowed
with the present carrier frequency. The remote interface also generates a "Data out of range" error.
l
Front Panel:
Then select a waveform shape.
l
SCPI:
FUNCtion {SINusoid|SQU|PULS|RAMP|TRI|ARB|NOISe|PRBS}
The
command configures a waveform with one command.
Carrier Frequency
The maximum carrier frequency varies by function, as shown below. The default is 1 kHz for all functions. Arbitrary
waveform "frequency" is also set using the
command.
Function
Minimum Frequency
Maximum Frequency
Sine
1 µHz
30 MHz
Square
1 µHz
30 MHz
Ramp
1 µHz
200 kHz
Pulse
1 µHz
30 MHz
PRBS
1 mBPS
50 mPBS
Noise BW
1 mHz
30 MHz
Arbitrary
1 µSa/s
250 MSa/s
l
Front Panel:
72
Agilent 33500 Series Operating and Service Guide
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......