Front-Panel Menu Operation Introduction
Output a Modulated Waveform
A modulated waveform consists of a carrier waveform and a modulating waveform. In AM (amplitude modulation), the
carrier amplitude is varied by the modulating waveform. For this example, you will output an AM waveform with 80%
modulation depth. The carrier will be a 5 kHz sine wave and the modulating waveform will be a 200 Hz sine wave.
1.
Select the function, frequency, and carrier amplitude.
Press the
Waveforms
key and then press
Sine
. Press the
Frequency
,
Amplitude
, and
Offset
softkeys to con-
figure the carrier waveform. For this example, select a 5 kHz sine wave with an amplitude of 5 Vpp, with 0 V offset.
Note that you may specify amplitude in Vpp, Vrms or dBm. To do this, either enter the value with the number pad
or press .
2.
Select AM.
Press
Modulate
and then select "AM" using the
Type
softkey. Then press
Modulate
to turn modulation on.
Notice that the
Modulate
button is illuminated, and the status message "AM Modulated by Sine" appears at the
top left of the display.
3.
Set the modulation depth.
Press the
AM Depth
softkey and then set the value to 80% using the numeric keypad or the knob and cursor
keys.
4.
Select the modulating waveform shape.
Press
Shape
to select the modulating waveform's shape. For this example, select a sine wave.
5. Press
More
and then
AM Freq
. Set the value to 200 Hz using the numeric keypad or the knob and cursor keys.
Press
Hz
to finish entering the number if you are using the numeric keypad.
Output an FSK Waveform
You can configure the instrument to "shift" its output frequency between two preset values using
. The
rate at which the output shifts between the two frequencies (called the "carrier frequency" and the "hop frequency") is
determined by the internal rate generator or the signal level on the rear-panel
Ext Trig
connector. For this example,
you will set the "carrier" frequency to 3 kHz and the "hop" frequency to 500 Hz, with an FSK rate of 100 Hz.
Agilent 33500 Series Operating and Service Guide
29
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......