APPLy Subsystem
APPLy Subsystem
The APPLy subsystem allows you to configure entire waveforms with one command. The general form of an APPLy com-
mand is shown below:
[SOURce[1|2]:]APPLy:<
function
> [<
frequency
> [,<
amplitude
> [,<
offset
>]]]
For example,
APPLy:SIN 1e4,1,0.1
replaces the following commands:
FUNCtion SIN
FREQ 1e4
VOLT 1
VOLT:OFF 0.1
OUTP ON
Not only is APPLy shorter, it avoids settings conflicts that occur when sending commands individually. In addition,
performs the following operations:
l
Sets trigger source to IMMediate (equivalent to
).
l
Turns off any modulation, sweep, or burst mode currently enabled and places the instrument in continuous wave-
form mode.
l
Turns on the channel output (
) without changing output termination setting (
).
l
Overrides the voltage autorange setting and enables autoranging (
).
The instrument can generate eight types of waveforms: DC voltage, gaussian noise, PRBS, pulse, ramp/triangle wave,
sine wave, square wave and arbitrary (user) waveform. Waveform-specific settings exist in the
.
You can also query current output configuration (
).
General Remarks
Amplitude
l
Changing amplitude may briefly disrupt output at certain voltages due to output attenuator switching. The ampli-
tude is controlled, however, so the output voltage will never exceed the current setting while switching ranges. To
prevent this disruption, disable voltage autoranging using
. The
command auto-
matically enables autoranging.
l
Limits Due to Output Termination: The offset range depends on the output termination setting. For example, if
you set offset to 100 mVDC and then change output termination from 50 Ω to "high impedance," the offset volt-
age displayed on the front panel doubles to 200 mVDC (no error is generated). If you change from "high imped-
ance" to 50 Ω, the displayed offset voltage will be halved. See
for details.
l
Limits Due to Unit Selection: The amplitude limits are determined by the output units selected.
l
You cannot specify output amplitude in dBm if output termination is set to high impedance.The units are auto-
matically converted to Vpp.
Agilent 33500 Series Operating and Service Guide
163