Self-Test Procedures
Error
Message and Meaning
Probable Cause
606
Self-test failed; waveform FPGA revision check failed
Processor tried to read the revision register from the waveform FPGA (U1005)
and received an invalid value, possibly due to un-programmed FPGA or internal
SPI communications failure.
607
Self-test failed
; waveform FPGA read back error
Processor was unable to write and read back from a test location in the waveform
FPGA (U1005), possibly due to un-programmed FPGA or internal SPI com-
munications failure.
608
Self-test failed; waveform FPGA security check failed
Waveform FPGA (U1005) failed internal security check, possibly due to FPGA fail-
ures (tests 605-608), invalid FPGA image, or malfunctioning security device
(U1007). Self-test exits on this failure.
609
Self-test failed; waveform FPGA security check failed
Waveform FPGA (U1005) failed internal security check, possibly due to FPGA fail-
ures (tests 605-608), invalid FPGA image, or malfunctioning security device
(U1007). Self-test exits on this failure.
610
Self-test failed; main PLL not locked
Waveform FPGA (U1005) unable to lock to internal 10MHz oscillator (U903 or
U905).
Main Board
611
Self-test failed
; FPGA PLL not locked
Waveform FPGA (U1005) unable to lock to internal sample clock generator IC
(U906).
Main Board
612
Self-test failed; Chan n, waveform memory PLL not locked
Waveform RAM for indicated channel (U1101 or U1102) was unable to lock to its
clock.
Main Board
Agilent 33500 Series Operating and Service Guide