[SOURce[1|2]:]AM:STATe {ON|1|OFF|0}[SOURce[1|2]:]AM:STATe?[SOURce[1|2]:]BPSK:STATe
[SOURce[1|2]:]AM:STATe {ON|1|OFF|0}
[SOURce[1|2]:]AM:STATe?
[SOURce[1|2]:]BPSK:STATe {ON|1|OFF|0}
[SOURce[1|2]:]BPSK:STATe
[SOURce[1|2]:]FM:STATe {ON|1|OFF|0}
[SOURce[1|2]:]FM:STATe
[SOURce[1|2]:]FSKey:STATe {ON|1|OFF|0}
[SOURce[1|2]:]FSKey:STATe
[SOURce[1|2]:]PM:STATe {ON|1|OFF|0}
[SOURce[1|2]:]PM:STATe
[SOURce[1|2]:]PWM:STATe {ON|1|OFF|0}
[SOURce[1|2]:]PWM:STATe?
Enables or disables modulation.
Parameter
Typical Return
{ON|1|OFF|0}, default OFF
0 (OFF) or 1 (ON)
Enable AM (could also substitute FM, BPSK, FSK, PM, or PWM):
AM:STAT ON
l
To avoid multiple waveform changes, enable modulation after configuring the other modulation parameters.
l
Only one modulation mode may be enabled at a time.
l
The instrument will not enable modulation with sweep or burst enabled. When you enable modulation, the sweep
or burst mode is turned off.
l
PWM is allowed only when pulse is the selected function.
See Also
Agilent 33500 Series Operating and Service Guide
175