Output Configuration
l
Front Panel:
or
l
SCPI:
OUTPut[1|2]:POLarity {NORMal|INVerted}
Sync Output Signal
A sync output is provided on the front-panel
Sync
connector. All of the standard output functions (except DC and
noise) have an associated
Sync
signal. For applications where you may not want to output the
Sync
signal, you can
disable the
Sync
connector. The
Sync
signal may be derived from either output channel in a two-channel instrument.
General Behavior
l
By default, the
Sync
signal is derived from channel 1 and is routed to the
Sync
connector (enabled).
l
When the
Sync
signal is disabled, the output level on the
Sync
connector is at a logic "low."
l
The polarity of the
Sync
signal is specified by
OUTPut:SYNC:POLarity {INVerted|NORMal}
.
l
Inverting a waveform (see
), does not invert the associated
Sync
signal .
l
For sine, pulse, ramp, square, and triangle waves, the
Sync
signal is a square wave that is "high" in the first half of
the cycle and "low" in the last half. The
Sync
signal’s voltages are TTL-compatible when its load impedance
exceeds 1 kΩ.
l
For arbitrary waveforms, the
Sync
signal rises at the beginning of the waveform and falls at the middle of the arbi-
trary waveform. You can override this default behavior by using
to specify the point within the arbi-
trary waveform at which the
Sync
signal transitions to "low."
Modulation
l
For internally-modulated AM, FM, PM, and PWM, the
Sync
signal is normally referenced to the modulating wave-
form (not the carrier) and is a square waveform with a 50% duty cycle. The
Sync
signal is a TTL "high" during the
first half of the modulating waveform. You can set up the
Sync
signal to follow the carrier waveform by using the
command
OUTPut:SYNC:MODE {CARRier|NORMal|MARKer}
when modulating with internal modulation.
l
For externally-modulated AM, FM, PM, and PWM, the
Sync
signal is referenced to the carrier waveform (not the
modulating waveform) and is a square waveform with a 50% duty cycle.
Agilent 33500 Series Operating and Service Guide
65
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......