Output Configuration
or
l
SCPI:
VOLTage:RANGe:AUTO {OFF|ON|ONCE}
The
command always enables autoranging.
Output Control
By default, channel output is disabled at power on to protect other equipment. To enable a channel's output, see below.
When channel output is enabled, the corresponding channel button is lit.
l
If an external circuit applies excessive voltage to a channel output connector, the instrument generates an error
message and disables the output. To re-enable output, remove the overload and turn the channel on again.
l
Front Panel:
l
l
SCPI:
The
command always enables the channel output connector.
Waveform Polarity
In normal mode (default), the waveform goes positive at the beginning of the cycle. Inverted mode does the opposite.
l
As shown below, the waveform is inverted relative to the offset voltage. The offset voltage remains unchanged
when the waveform is inverted.
l
The
Sync
signal associated with an inverted waveform is not inverted.
64
Agilent 33500 Series Operating and Service Guide