-8 dB Range Flatness Adjustment
This completes the adjustment procedures for the one-channel instrument. Verification of the output specifications is
recommended.
If you are making adjustments to a two-channel instrument, continue with the next procedure in this section.
Agilent 33500 Series Operating and Service Guide
397