Set Up Arbitrary Waveform
2. Change the
Cycle Period
to 2.04 seconds and then press the
Sampling Rate
softkey. It will now be set to 200
Sa/s in order to play the 408 point waveform in 2.04 seconds.
3. Press
Done
to exit the parameter editing screen. Press
Edit Points
and notice that the
Point #
softkey is high-
lighted. Enter the number 160 and press
Enter
to move the marker.
4. Press
Voltage
and change the voltage of the selected point to 4.2 V. Press
Point #
and change the point marker
to 150 to move the marker off the point. When you press
Enter
to finish entering point 150, you will see the 4.2 V
anomaly in the wave that you just created at point 160.
Zoom and Pan
1. To see the point in detail, press
System
, then
Pan/Zoom Control
. Notice that the first softkey is set to
Hor-
izontal
, meaning that the zooming that we are about to do will be along the horizontal (time) axis. Change the
Zoom
to 500%, and the sine wave anomaly will be more obvious.
Agilent 33500 Series Operating and Service Guide
49