Frequency Sweep
l
You cannot configure the marker frequency with the front panel menus unless the
Sync
source is the sweeping
channel.
l
Front Panel:
l
SCPI:
[SOURce[1|2]:]MARKer:FREQuency {<freq>|MIN|MAX}
Sweep Trigger Source
In sweep mode, the instrument outputs a single sweep when a trigger signal is received. After one sweep from the
start frequency to the stop frequency, the instrument waits for the next trigger while outputting the start frequency.
l
Sweep trigger source: Internal (default), External, Time, or Manual.
l
With the Internal (immediate) source, the instrument outputs a continuous sweep at a rate determined by the
total of the hold time, sweep time and return time. The sweep time for this source is limited to 8000 seconds.
l
With the External source, the instrument accepts a hardware trigger on the rear-panel Ext Trig connector and ini-
tiates one sweep each time Ext Trig receives a TTL pulse with the specified polarity.
l
The trigger period must be greater than or equal to the specified sweep time.
l
With the Manual source, the instrument outputs one sweep each time the front-panel
Trigger
key is pressed.
l
Front-Panel:
l
To specify the slope of the trigger signal edge:
l
SCPI:
TRIGger[1|2]:SOURce {IMM|EXT|TIMer|BUS}
Agilent 33500 Series Operating and Service Guide
95
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......