IEEE-488 Common Commands
*ESE <
enable_value
>
*ESE?
Event Status Enable Command and Query
. Enables bits in the
for the
group.
The selected bits are then reported to bit 5 of the Status Byte Register.
Parameter
Typical
Return
Decimal sum of the bits in the register (table below), default 0. For example, to enable bit 2 (value
4), bit 3 (value 8), and bit 7 (value 128), the decimal sum would be 140 (4 + 8 + 128). Default 0.
+48
Enable bit 4 (value 16) and bit 5 (value 32) in the enable register:
*ESE 48
l
Use
to control whether the Standard Event enable register is cleared at power on. For example,
pre-
serves the enable register contents through power cycles.
l
does not clear enable register, does clear event register.
*ESR?
Standard Event Status Register Query
. Queries the event register for the
group. Register is
read-only; bits not cleared when read.
Parameter
Typical Return
(none)
+24
Read the event register (bits 3 and 4 are set).
*ESR
l
Any or all conditions can be reported to the Standard Event summary bit through the enable register. To set the
enable register mask, write a decimal value to the register using
l
Once a bit is set, it remains set until cleared by this query or
*IDN?
Identification Query
. Returns instrument’s identification string.
Parameter
Typical Return
(none)
Agilent Technologies,33522B,XXXXXXXXXX,0.179-1.19-8.88-52-00
Return the instrument's identification string:
*IDN?
l
Identification string contains four comma separated fields:
l
Manufacturer name
l
Model number
Agilent 33500 Series Operating and Service Guide