Ratio Error
Example
This example demonstrates how to compute the total measurement error
when using the Ratio Function. Assume a ratio measurement is made
with the following conditions:
•
Channel 1
, 5 V input on the 10 Volt range
•
Channel 2
, 10 mV input on the 10 mV range
To compute the total measurement error using the 90-day accuracy
specifications:
Ratio Error % = Channel 1 accuracy in % + Channel 2 accuracy in %
Channel
1
Reading
Error
%
=
0.0020%
Channel
1
Range
Error
%
=
0.0004%
∗
Range
Rea ding
=
.0004%
∗
10
5
=
.0008%
Total
Channel
1
Error
%
=
0.0020%
+
0.0008%
=
.0028%
Channel
2
Reading
Error
%
=
0.0040%
Channel
2
Range
Error
%
=
0.0002%
∗
Range
Reading
=
.0002%
∗
10
mV
10
mV
=
.0002%
Total
Channel
2
Error
%
=
0.0040%
+
0.0002%
=
.0042%
Ratio
Error
%
=
0.0028%
+
0.0042%
=
.0070%
Ratio
=
5
V
10
mV
=
500
±
0.0070%
=
500
±
0.035
For Ratio Measurements, if both input channels use the same range, use
0 for the % of reading error.
Chapter 8
To Calculate Total Measurement Error
284
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......