v
10 Move down a level to view the first stored reading.
Reading memory is automatically turned off when you go to the
“parameter” level in the menu.
The first reading displayed is the first reading that was stored (FIFO). If no
readings are stored in memory, “EMPTY” is displayed. The first five
digits of the reading and the reading number are displayed. After a brief
pause, the reading number is removed and all reading digits are shown.
The stored readings are displayed with their units (
µ
, m, k, etc.) when
appropriate. For example:
Reading
number
10.31607K: 11
Units
10.316076 K
> >
11 Move across to view the two remaining stored readings.
The readings are stored
horizontally
on the “parameter” level.
If you press
<
when you first get to the “parameter” level, you will see the
last reading and you will know how many readings were stored.
On/Off
Shift
<
12 Turn off the menu.
EXITING
Chapter 2 Front-Panel Operation
To Use Reading Memory
52
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......