![Agilent Technologies 34420A Скачать руководство пользователя страница 169](http://html1.mh-extra.com/html/agilent-technologies/34420a/34420a_user-manual_2868276169.webp)
The Questionable Data Register
The
questionable data
register provides information about the quality of the
meter’s measurement results. Overload conditions can be reported in the
questionable data summary bit through the enable register. You must write a
decimal value using the
STATus:QUEStionable:ENABle
command to set
the enable register mask.
Note:
A reading overload condition is always reported in both the
standard event register (bit 3) and the questionable data event register
(bits 0, 1, or 9). However, no error message is recorded in the meter’s error
queue.
Bit Definitions — Questionable Data Register
Bit
Decimal
Value
Definition
0
Voltage Overload
1
Range overload on dc volts
1
Not Used
2
Always set to 0.
2
Not Used
4
Always set to 0.
3
Not Used
8
Always set to 0.
4
Temperature Overload
16
Range overload on temperature
measurements.
5
Not Used
32
Always set to 0.
6
Not Used
64
Always set to 0.
7
Not Used
128
Always set to 0.
8
Not Used
256
Always set to 0.
9
Ohms Overload
512
Range overload on 2-wire or 4-wire
Ohms
10
Not Used
1024
Always set to 0.
11
Not Used
2048
Always set to 0.
12
Not Used
4096
Always set to 0.
13
Not Used
8192
Always set to 0.
14
Not Used
16384
Always set to 0.
15
Not Used
32768
Always set to 0.
4
Chapter 4 Remote Interface Reference
The SCPI Status Model
167
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......