To Secure Against Calibration
You can secure the meter against
calibration either from the front panel or remote interface.
The meter is
secured when shipped from the factory, and the security code is set to
“HP034420”
.
Be sure to read the security code rules on page 107 before attempting to
secure the meter.
•
Front-panel operation:
1: UNSECURED
If the meter is unsecured, you will see the above command when you
go into the CAL MENU. To secure the meter, select the “parameter”
level of the UNSECURED command, enter the security code, then
press Menu Enter.
∧
000000 CODE
•
When you go to the “commands” level in the CAL MENU again, you
will notice that the meter is secured. Notice also that the
2: CALIBRATE
command is now hidden and you cannot perform a
calibration.
1: UNSECURED
•
Remote interface operation:
CALibration:SECure:STATe {OFF|ON},<
code
>
To secure the meter, send the above command with the same code as
used to unsecure. For example,
CAL:SEC:STAT ON, HP034420
3
Chapter 3 Features and Functions
Calibration
109
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......