![Agilent Technologies 34420A Скачать руководство пользователя страница 284](http://html1.mh-extra.com/html/agilent-technologies/34420a/34420a_user-manual_2868276284.webp)
Interpreting Meter Specifications
This section is provided to give you a better understanding of the
terminology used and will help you interpret the meter’s specifications.
Number of Digits and Overrange
The
number of digits
specification is the most fundamental, and
sometimes, the most confusing characteristic of a meter. The number of
digits is equal to the maximum number of "9’s" the meter can measure or
display. This indicates the number of full digits. Most meters have the
ability to overrange and add a partial or "1/2" digit.
For example, the 34420A can measure 9.999999 Vdc on
the 10 V range. This represents seven full digits of resolution. The
meter can also overrange on the 10 V range and measure up to a
maximum of 12.000000 V. This corresponds to a 7 1/2-digit measurement
with 20% overrange capability.
Sensitivity
Sensitivity is the minimum level that the meter can detect for a given
measurement. Sensitivity defines the ability of the meter to respond to
small changes in the input level. For example, suppose you are
monitoring a 1 mV signal and you want to adjust the level to
within
±
1
µ
V. To be able to respond to an adjustment this small, this
measurement would require a meter with a sensitivity of at least 1
µ
V.
You could use a 6 1/2-digit meter if it has a 1 Vdc or smaller range. You
could also use a 4 1/2-digit meter with a 10 mV range.
8
Chapter 8
Interpreting Meter Specifications
285
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......