To Set the Baud Rate
You can select one of six baud rates for
RS-232
operation. The rate is set
to 9600 baud when the meter is shipped from the factory.
See also “Baud
On/Off
Shift
<
1
Turn on the front-panel menu.
A: MEAS MENU
<
<
2
Move across to the
I/O MENU
choice on this level.
F: I/O MENU
∨
> >
3
Move down a level and then across to the
BAUD RATE
command.
3: BAUD RATE
∨
4
Move down to the “parameter” level to select the baud rate.
Use the left/right arrow keys to see the baud rate choices. Choose from
one of the following: 300, 600, 1200, 2400, 4800, or
9600
baud.
9600 BAUD
Auto/Man
5
Save the change and exit the menu.
ENTER
The baud rate selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
Remote Interface:
you can set or query the baud rate using the
following commands:
SYStem:COMMunicate:SERial
:BAUD <
baud>
:BAUD?
Chapter 4 Remote Interface Reference
To Set the Baud Rate
190
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......