To Set the Integration Time
Integration time is specified in Number of Power Line Cycles (NPLC).
You can set the NPLC to 0.02, 0.2, 1, 2, 10, 20, 100, or 200.
•
You can set the integration time to one of three fixed values by
choosing the number of digits displayed,
•
You can set the integration time in the MEASure menu using the
INTEGRATE command.
•
The Integration Time is directly related to the maximum number of
digits the meter will display.
NPLC
Filter Off
Max Digits Shown
Filter On
Max Digits Shown
0.02
4
1
⁄
2
5
1
⁄
2
0.2
5
1
⁄
2
6
1
⁄
2
1
6
1
⁄
2
7
1
⁄
2
2
6
1
⁄
2
7
1
⁄
2
10
6
1
⁄
2
7
1
⁄
2
20
7
1
⁄
2
7
1
⁄
2
100
7
1
⁄
2
7
1
⁄
2
200
7
1
⁄
2
7
1
⁄
2
•
You can always show fewer digits than the maximum allowed
(the minimum number of digits shown is 4
1
⁄
2
).
Integration Time is local to the selected function. This means that you
can select the integration time for each function independently. The
meter remembers integration time when you switch between functions.
1
Chapter 1 Quick Start
To Set the Integration Time
25
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......