Using the READ? Command
The
READ?
command changes the state of the trigger system from the
“idle” state to the “wait-for-trigger” state. Measurements will begin when
the specified trigger conditions are satisfied following the receipt of the
READ?
command. Readings are sent
immediately
to the output buffer.
You
must
enter the reading data into your bus controller or the meter
will stop making measurements when the output buffer fills. Readings
are not
stored in the meter’s internal memory when using the
READ?
command.
Sending the READ? command is like sending the INITiate command
followed immediately by the FETCh? command, except readings are not
buffered internally.
C a u t i o n
If you send two query commands without reading the response from the
first, and then attempt to read the second response, you may receive some
data from the first response followed by the complete second response. To
avoid this, do not send a query command without reading the response.
When you cannot avoid this situation, send a device clear before sending
the second query command.
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
130
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......