Triggering and Memory
Reading HOLD Sensitivity 10%, 1%, 0.1%, or 0.01% of range
Samples/Trigger
1 to 50,000
Trigger Delay
0 to 3600s; 10
µ
s step size
External Trigger Delay
<1ms
External Trigger Jitter
<500
µ
s
Memory
1024 readings
Math Functions
NULL (Channel 1 DCV, Channel 2 DCV, Difference,
MMMN
Resistance, Temperature)
STATS (Min, Max, Average, Peak-Peak, Standard Deviation,
MMMMII
Number of readings)
SCALE (Allows linear scaling as y=mx+b)
CHART NULL (Establishes zero for rear panel output)
Filter (Analog or Digital or Both)
Analog
Low pass 2 pole @ 11Hz, available
for 1, 10, 100mV range
Digital
Moving average filter, 10 (fast),
50 (medium) or, 100 (slow) reading
averages.
Chart Out (Analog Out)
Maximum output:
±
3V
Accuracy:
±
0.1% of 1 mV
Output Resistance: 1 k
Ω
±
5%
Update rate: once per reading
Span and Offset: Adjustable
Standard Programming Languages
SCPI (IEEE 488.2), Keithley 181
Accessories Included
4 ft low thermal cable with copper spade lugs, 4-wire shorting
plug, user’s manual, service manual, contact cleaner, test
report and power cord
General Specifications
Front Panel Connection shielded, low thermal, 99% copper
contacts
Power Supply
100V/120V/220V(230V)/240V
±
10%
Power Line Frequency
45Hz to 66Hz and 360Hz to 440Hz
Automatically sensed at power-on.
Power Consumption
25VA peak (10W average)
Operating Environment Full accuracy for 0
°
C to 55
°
C
Full accuracy to 80% R.H. up to 40
°
C
Storage Environment
-40
°
C to 75
°
C
Weight
3 kg (6.5 lbs)
Safety
Designed to CSA, UL-1244,
IEC-1010
RFI and ESD
CISPR 11
Vibration and Shock
MIL-T-28800E, Type III, Class 5
Warranty
3 year
8
Chapter 8 Specifications
34420A Nano Volt/Micro Ohm Meter
279
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......