The Status Byte
The status byte
summary register
reports conditions from other status
registers. Query data that is waiting in the meter’s output buffer is
immediately reported through the “message available” bit (bit 4). Bits in
the summary registers are
not
latched. Clearing an event register will
clear the corresponding bits in the status byte summary register. Reading
all messages in the output buffer, including any pending queries, will
clear the message available bit.
Bit Definitions — Status Byte Register
Bit
Decimal
Value
Definition
0
Not Used
1
Always set to 0
1
Not Used
2
Always set to 0
2
Not Used
4
Always set to 0
3
Questionable Data
8
One or more bits are set in the Questionable
Data register (bits must be "enabled" in the
enable register).
4
Message Available
16
Data is available in the meter output buffer
5
Standard Event
32
One or more bits are set in the Standard
Event register (bits must be "enabled" in the
enable register).
6
Request Service
64
The meter is requesting service (serial poll).
7
Operational Status
128
One or more bits are set in the Operational
Event register (bits must be "enabled" in the
enable register)
4
Chapter 4 Remote Interface Reference
The SCPI Status Model
161
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......