Loading Errors Due to Input Resistance (dc volts)
Measurement loading errors occur when the resistance of the device-
under-test (
DUT
) is an appreciable percentage of the meter’s own input
resistance. The diagram below shows this error source.
V
s
=
Ideal DUT voltage
R
S
=
DUT source resistance
R
i
=
Meter input resistance
(10 M
Ω
or > 10 G
Ω
)
Error
(
%
)
=
100
x R
S
R
S
+
R
i
Ideal
Meter
7
Chapter 7 Measurement Tutorial
Measurement Techniques and Sources of Error