To Use Reading Hold
The reading hold feature allows you to capture and hold a stable reading
on the display. When a stable reading is detected, the meter emits a beep
and holds the value on the display.
•
Reading hold has an adjustable
sensitivity band
to allow you to select
which readings are considered stable enough to be displayed. The
band is expressed as a percent of reading on the selected range. The
meter will capture and display a new value only after
three
consecutive readings are within the band.
•
The default band is 0.10% of reading. After enabling reading hold,
you can choose a different band by pressing
Shift >
(Menu Recall).
This takes you to the “READ HOLD” command in the TRIG MENU
(
only if reading hold is enabled
).
Go down to the “parameter” level, and then select a value:
0.01%,
0.10%
, 1.00%, or 10.00% of reading.
•
The sensitivity band is stored in
volatile
memory; the meter sets the
band to 0.10% of reading when power has been off or after a remote
interface reset.
Toggles between autotrigger
and reading hold
*
(sample) annunciator is on
during each measurement
Chapter 2 Front-Panel Operation
To Use Reading Hold
48
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......