Using
“ ? ”
Commands
The bus controller may send commands at any
time, but a
SCPI
instrument may only send responses when
specifically
instructed to do so. Only query commands (commands that end with
a “?”) will instruct the instrument to send a response message. Queries
return either measured values or internal instrument settings.
C a u t i o n
If you send two query commands without reading the response from the
first, and then attempt to read the second response, you may receive some
data from the first response followed by the complete second response. To
avoid this, do not send a query command without reading the response.
When you cannot avoid this situation, send a device clear before sending
the second query command.
Using
“ * ”
Commands
Commands starting with a “ * ” are called
common commands
. They are required to perform the identical function
for
all
instruments that are compliant with the
IEEE-488.2
interface
standard. The “ * ” commands are used to control reset, self-test, and
status operations in the meter.
Chapter 4 Remote Interface Reference
An Introduction to the SCPI Language
184
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......