Automatic Trigger Delays
If you do not specify a trigger delay, the meter selects an automatic delay
for you. The delay is determined by function, range, and integration time.
•
DC voltage :
Range
NPLC
<
1
Trigger Delay
NPLC
³
1
Trigger Delay
1
mV
10
mV to 120 V
15 ms
1.0 ms
15 ms
1.5 ms
•
Resistance (2-wire and 4-wire):
Range
NPLC < 1
Trigger Delay
NPLC
³
1
Trigger Delay
1
Ω
10
Ω
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
1.0 ms
1.0 ms
1.0 ms
1.0 ms
1.0 ms
4 ms
40 ms
1.5 ms
1.5 ms
1.5 ms
1.5 ms
1.5 ms
6 ms
60 ms
•
Temperature:
NPLC < 1
Trigger Delay
NPLC
³
1
Trigger Delay
RTD
THERmistor
ThermoCouple
1.0 ms
1.0 ms
1.0 ms
1.5 ms
1.5 ms
1.5 ms
3
Chapter 3 Features and Functions
Triggering
89