![Agilent Technologies 34420A Скачать руководство пользователя страница 205](http://html1.mh-extra.com/html/agilent-technologies/34420a/34420a_user-manual_2868276205.webp)
-221
Settings conflict
This error can be generated in one of the following situations:
You sent a
CONFigure
or
MEASure
command with autorange enabled
and with a fixed resolution. Example:
CONF:VOLT:DC DEF,0.1
You turned math on (
CALC:STAT ON
) and then changed to a math
operation that was not valid with the present measurement function.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example:
TRIG:COUN -3
-223
Too much data
A character string was received but could not be executed because the
string length was more than 11 characters. This error can be generated
by the
CALibration:STRing
and
DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the
command. You may have used an invalid parameter choice. Example:
CALC:FUNC SCALE
(
SCALE
is not a valid choice).
-230
Data corrupt or stale
A
FETCh?
command was received but internal reading memory was
empty. The reading retrieved may be invalid.
-330
Self-test failed
The meter’s complete self-test failed from the remote interface (
*TST?
command). In addition to this error, more specific self-test errors are also
reported.
See also “Self-Test Errors,” starting on page 208.
Chapter 5 Error Messages
Execution Errors
204
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......