![Agilent Technologies 34420A Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/agilent-technologies/34420a/34420a_user-manual_2868276077.webp)
Statistics
•
After you enable statistics, the first reading that the meter takes is
stored as both the minimum and maximum value. The
minimum
is
replaced with any subsequent value that is less. The
maximum
is
replaced with any subsequent value that is greater.
•
All values are stored in
volatile
memory; the meter clears the values
when stats is turned on, when power has been off, or after a remote
interface reset or preset.
•
When enabled, the following statistics are available:
MIN
minimum reading
MAX
maximum reading
AVER
arithmetic average of all readings
STD DEV
standard deviation of all readings
PEAK – PEAK peak to peak value of minimum and maximum
COUNT
the total number of readings
•
Front Panel operation:
Enable the statistics by pressing
the
Stats
key. Read the statistics gathered in the MATH menu
under the STATS command.
After reading the stored statistics, press
Shift <
(Menu On/Off) to
return to the measurement and continue taking statistics.
3
Chapter 3 Features and Functions
Math Operations
75
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......