To Select the Programming Language
You can select one of three language to program the meter from the
selected remote interface. The language is
SCPI
when the meter is
shipped from the factory. The languge setting is stored in
non-volatile
memory and does not change with power off or reset.
See also
“Programming Language Selection,” on page 105.
On/Off
Shift
<
1
Turn on the front-panel menu.
A: MEAS MENU
<
<
2
Move across to the
I/O MENU
choice on this level.
F: I/O MENU
∨
<
3
Move down a level and then across to the
LANGUAGE
command.
5: LANGUAGE
∨
4
Move down to the “parameter” level to select the language.
Choose from one of the following:
SCPI
or KEITHLEY:181
SCPI
Auto/Man
5
Save the change and turn off the menu.
ENTER
The language selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
Remote Interface:
you can set the programming language using the
following commands:
SYStem:LANGuage {SCPI | 181}
L1X
(SCPI)
L2X
(Keithley 181)
Chapter 4 Remote Interface Reference
To Select the Programming Language
192
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......