DISASSEMBLY OF THE RATCHET HANDLE FOR FLASH
STERILIZATION
The ratchet handle is a three-piece assembly which can be disassembled to facilitate easy cleaning and maintenance.
•
Grasp the handle at the knurled end with the slot in the shaft below the ratchet facing you. Take a blunt
instrument and insert it in the indentation within the slot on the shaft (See Fig. 13) and pull downward toward the
knurled handle. This will release the ratchet, and it should fall out of the assembly. This will expose the next place
to disassemble. (See Fig. 14.)
•
Reach into the exposed hole, grasp the pin, and pull it out of the assembly. This completes the disassembly.
(See Fig. 15.)
•
Reassemble the instrument by reversing the above procedure.
STORAGE CONDITIONS
The system should be stored under normal warehouse conditions.
ACCESSORIES
Description:
Meshgraft
II Tissue Expansion System Complete (includes 00-2195-001-00, 00-2195-006-
00, 00-2195-022-00)
REF:
00-2195-000-00
Description:
Meshgraft
II Tissue Expansion System Instrument
REF:
00-2195-001-00
Description:
Meshgraft
II Autoclave Case
REF:
00-2195-006-00
Description:
Dermacarrier
II Skin Graft Carriers (Sold Separately)
REF:
00-2195-012-00
1.5:1 Expansion Ratio
Box of 20
00-2195-013-00
3:1 Expansion Ratio
Box of 20
00-2195-014-00
6:1 Expansion Ratio
Box of 20
00-2195-015-00
9:1 Expansion Ratio
Box of 20
Description:
Meshgraft
II Tissue Expansion System Ratchet Handle
REF:
00-2195-022-00
SERVICE INFORMATION
The
Meshgraft
II Tissue Expansion System should be returned to Zimmer Surgical, Dover, OH for servicing.
NOTE: Zimmer cannot be held liable for any instrument malfunction resulting from repairs or service performed by an
unauthorized service center. Do not attempt to disassemble the unit. It is a factory sealed unit with no user serviceable
parts inside.
The
Meshgraft
II Tissue Expansion System should be returned every 12 months for inspection and preventive
maintenance. Annual factory calibration checks are strongly recommended to verify continued accuracy.
RETURN AUTHORIZATION AND REPLACEMENT INFORMATION
When it is necessary to return the instrument for inspection and preventive maintenance or repair; within the U.S.A. call
1-800-830-0970 to receive a Return Goods Authorization (RGA) number. Outside the U.S.A., please contact your local
Zimmer representative.
The instrument must be properly packaged when sent into the factory. If the original packaging is no longer available,
proper packaging can be requested when the Return Goods Authorization (RGA) number is received.
A purchase order must accompany all equipment for repair. The customer will be responsible for all shipping charges.
The expected longevity of the
Meshgraft
II Tissue Expansion System is ten years. Zimmer recommends that units over
this age should be replaced because they have exceeded their normal useful life. Please contact your Zimmer sales
representative regarding replacement options.
LIMITED WARRANTY (U.S.A. only)
Zimmer Surgical warrants that the
Meshgraft
II Tissue Expansion System, along with all parts and accessories, has
been tested and inspected, and has left the factory in proper working condition, free from visible defects.
Zimmer Surgical warrants to the first consumer purchaser of new
Meshgraft
II Tissue Expansion Systems, Ratchet
Handles, and Autoclave Cases that these products and accessories will, under normal and reasonable use, be free from
defects in material and workmanship for one (1) year after the date of shipment from the factory. During the warranty
period, Zimmer Surgical shall repair (or at its sole option replace) the defective product or part without cost to the
purchaser. Defective parts replaced under this warranty shall become the property of Zimmer Surgical. This limited
warranty does not cover damage caused by misuse, abuse, accident, neglect, ordinary wear and tear, or any uses not
prescribed in this manual. If the unit becomes defective because of misuse or abnormal conditions of operation, repairs
will be billed at our current rate.
Summary of Contents for Meshgraft II
Page 95: ...95 1 ...
Page 97: ...97 3 ...
Page 101: ...101 7 ...
Page 102: ...102 8 ...
Page 103: ...103 9 ...
Page 104: ...104 10 ...
Page 105: ...105 11 ...
Page 106: ...106 12 ...
Page 107: ...107 13 ...
Page 108: ...108 14 ...
Page 109: ...109 15 ...
Page 111: ...111 ...