IM 11G06A01-02EN
5-41
5. SCREENS
1st Edition : Oct. 27, 2023-00
(6) If you operated the valve yourself, return it to its original condition here.
• To continue calibration, touch the [Continue]. The Calibration menu
screen will be displayed.
• To exit, tap [End ].
The replacement time screen will appear. (The replacement time is the
same as the replacement time for auto-calibration. See 5.1.2 Auto
If the replacement time has elapsed when the "End" button is touched, the
measurement screen will be displayed.
[Continue]
[End]
● In the replacement time in progress screen, all Ch (components) as
well as the calibration target are displayed. The display range is shown
according to the range-switching setting.
Make sure that the influence of the calibration gas no longer exists.
After the replacement time has elapsed, the measurement screen is
automatically displayed.
Tapping the " Abort" immediately terminates the replacement time and
displays the measurement screen.
● Contact output during manual span calibration
When "Ch○ for span calibration" (○: 1 to 5), "for sample switching" or "system calibration" is
assigned to the contact output, the contact output changes as shown in Figure 5.11 and Figure
5.12 according to the gas flow channel setting in the auto-calibration setting. (See 5.2.3 Contact
output setting for the function assigned to the contact output and 5.1.2 Auto Calibration Setting
for the gas flow channel.)
for sample
switching
for CHx
span calibration Tap [Next].
End of calibration
Tap [Enter].
Figure 5.11
Gas flow channel setting: Analyzer
for system
calibration
for CHx span
calibration Tap[Nest].
End of calibration
Tap [Enter].
The contact output for "sample switching" remains ON.
Figure 5.12
Gas flow channel setting: Sampling system